Human NT-ProANP Quantikine ELISA Kit
Human NT-ProANP Quantikine ELISA Kit
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human NT-ProANP Quantikine ELISA Kit
Catalog #
|
Availability
|
Size / Price
|
DANP00
|
Contact your distributor for availability.
|
1 Kit /
|
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (25 uL), EDTA Plasma (25 uL), Heparin Plasma (25 uL)
Assay Range
0.3 - 20 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human NT-ProANP.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine® Human NT-ProANP Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human NT-ProANP in cell culture supernates, serum, and plasma. It contains E. coli-expressed recombinant human NT-ProANP and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human NT-ProANP showed linear curves that were parallel to the standard curves obtained using the Quantikine® kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring human NT-ProANP.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
1.71
|
6.04
|
11.1
|
1.81
|
5.95
|
11.5
|
Standard Deviation
|
0.069
|
0.224
|
0.583
|
0.129
|
0.426
|
0.842
|
CV%
|
4
|
3.7
|
5.3
|
7.1
|
7.2
|
7.3
|
Recovery
The recovery of human NT-ProANP spiked to levels throughout the range of the assay was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=4)
|
104
|
87-114
|
EDTA Plasma (n=4)
|
102
|
84-120
|
Heparin Plasma (n=4)
|
98
|
84-111
|
Serum (n=4)
|
108
|
95-122
|
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of human NT-ProANP were diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.