全国统一客服热线:
0755-28715175

产品中心
Product
4产品中心
Human CXCL11/I-TAC Quantikine ELISA Kit
Human CXCL11/I-TAC Quantikine ELISA Kit
Human CXCL11/I-TAC Quantikine ELISA Kit

Human CXCL11/I-TAC Quantikine ELISA Kit

 详情说明

Human CXCL11/I-TAC Quantikine ELISA Kit Summary

货号:DCX110  规格:96T

深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!

Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL)
Sensitivity
39.7 pg/mL
Assay Range
62.5 - 4,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human I-TAC
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.

Product Summary

The Quantikine Human I-TAC Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human I-TAC in cell culture supernates, serum, and plasma. It contains E. coli-expressed recombinant human I-TAC and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human I-TAC showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring human I-TAC.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision

Serum, EDTA Plasma, Heparin Plasma

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 40 40 40
Mean 403 1321 2481 437 1403 2644
Standard Deviation 19 62.5 147.3 35.1 107 190
CV% 4.7 4.7 5.9 8 7.6 7.2

Cell Culture Supernates

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 40 40 40
Mean 353 1179 2191 409 1290 2401
Standard Deviation 18 40.6 102.9 35.5 93 179
CV% 5.1 3.4 4.7 8.7 7.2 7.5

Recovery

The recovery of I-TAC spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 100 91-111
EDTA Plasma (n=5) 98 85-112
Heparin Plasma (n=5) 98 90-106
Serum (n=5) 106 93-117

Linearity

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of I-TAC were serially diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Human CXCL11/I-TAC ELISA Linearity

Copyright © 深圳9游总区生物科技有限公司版权所有

扫一扫,加关注

客户服务热线:
0755-33164177
13924667705
18665335907
0755-28715175


线