Human GM-CSF Quantikine ELISA Kit
Human GM-CSF Quantikine ELISA Kit
Human GM-CSF Quantikine ELISA Kit Summary
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DGM00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3.5 hours or 4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Citrate Plasma (100 uL)
Assay Range
7.8 - 500 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human GM-CSF
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human GM-CSF Immunoassay is a 3.5 or 4.5 hour solid phase ELISA designed to measure GM-CSF levels in cell culture supernates, serum, and plasma. It contains E. coli-expressed recombinant human GM-CSF and antibodies raised against the recombinant factor. It has been shown to quantitate recombinant human GM-CSF accurately. Results obtained using natural human GM-CSF showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural GM-CSF.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
40
|
40
|
40
|
Mean
|
54.7
|
156
|
321
|
53.1
|
157
|
317
|
Standard Deviation
|
1.1
|
3.2
|
4.8
|
3.3
|
9.1
|
15.4
|
CV%
|
2
|
2.1
|
1.5
|
6.2
|
5.8
|
4.9
|
Cell Culture Supernates
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
40
|
40
|
40
|
Mean
|
46.8
|
131
|
269
|
45.1
|
133
|
266
|
Standard Deviation
|
1.1
|
3.6
|
7.3
|
2.4
|
7
|
11.5
|
CV%
|
2.4
|
2.7
|
2.7
|
5.3
|
5.3
|
4.3
|
Recovery
The recovery of GM-CSF spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=5)
|
97
|
85-104
|
Citrate Plasma (n=5)
|
101
|
92-106
|
EDTA Plasma (n=5)
|
100
|
94-105
|
Heparin Plasma (n=5)
|
98
|
92-104
|
Serum (n=5)
|
98
|
93-103
|
Linearity
To assess the linearity of the assay, five samples were spiked with high concentrations of GM-CSF in various matrices and diluted with the appropriate Calibrator Diluent to produce values within the dynamic range of the assay.