Human ICAM-1/CD54 Allele-specific Quantikine ELISA Kit
Human ICAM-1/CD54 Allele-specific Quantikine ELISA Kit
Human ICAM-1/CD54 Allele-specific Quantikine ELISA Kit Summary
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DCD540
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL)
Assay Range
1.6 - 50 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human sICAM-1
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human sICAM-1 immunoassay is a 2.0 hour solid phase ELISA designed to measure sICAM-1 in cell culture supernates, serum, and plasma. It contains NS0-expressed recombinant human sICAM-1 and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human sICAM-1 showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring sICAM-1.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
40
|
40
|
40
|
Mean
|
4.61
|
12.1
|
19.6
|
4.77
|
12.5
|
20.5
|
Standard Deviation
|
0.17
|
0.63
|
0.98
|
0.32
|
0.67
|
0.9
|
CV%
|
3.7
|
5.2
|
5
|
6.7
|
5.4
|
4.4
|
Recovery
The recovery of sICAM-1 spiked to levels throughout the range of the assay was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=4)
|
104
|
90-109
|
Linearity
To assess the linearity of the assay, samples spiked with or containing high concentrations of sICAM-1 were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.