Human G-CSF Quantikine ELISA Kit Summary
Human G-CSF Quantikine ELISA Kit Summary
Human G-CSF Quantikine ELISA Kit Summary
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DCS50
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3.5 hours or 4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Citrate Plasma (100 uL)
Assay Range
39.0 - 2,500 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human G-CSF
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human G-CSF Immunoassay is a 3.5 or 4.5 hour solid phase ELISA designed to measure G-CSF in cell culture supernates, serum, and plasma. It contains E. coli-expressed recombinant human G-CSF and antibodies raised against the protein. It has been shown to accurately quantitate recombinant human G-CSF. Results obtained using natural human G-CSF showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural human G-CSF.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
177
|
515
|
1047
|
143
|
818
|
1514
|
Standard Deviation
|
4.5
|
9.4
|
22
|
17
|
72
|
77.1
|
CV%
|
2.5
|
1.8
|
2.1
|
11.9
|
8.8
|
5.1
|
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
280
|
827
|
1696
|
176
|
1094
|
2169
|
Standard Deviation
|
7.8
|
14.1
|
19.5
|
7.3
|
34.6
|
81.6
|
CV%
|
2.8
|
1.7
|
1.1
|
4.1
|
3.2
|
3.8
|
Recovery
The recovery of G-CSF spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=5)
|
94
|
85-110
|
Citrate Plasma (n=5)
|
90
|
78-111
|
EDTA Plasma (n=5)
|
93
|
80-117
|
Heparin Plasma (n=5)
|
87
|
76-103
|
Serum (n=5)
|
88
|
75-106
|
Linearity
To assess the linearity of the assay, five samples were spiked with high concentrations of G-CSF in various matrices and diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.