Human IL-17 Quantikine ELISA Kit Summary
Human IL-17 Quantikine ELISA Kit Summary
Human IL-17 Quantikine ELISA Kit Summary
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:D1700
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Citrate Plasma (100 uL)
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Natural and recombinant human IL-17
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human IL-17 Immunoassay is a 4.5 hour solid phase ELISA designed to measure human IL-17 in cell culture supernates, serum, and plasma. It contains E. coli-expressed recombinant human IL-17 and antibodies raised against the recombinant factor. This immunoassay has been shown to quantitate recombinant human IL-17 accurately. Results obtained using natural human IL-17 showed linear curves that were parallel to the standard curves obtained using the recombinant Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural human IL-17.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
131
|
736
|
1480
|
125
|
408
|
794
|
Standard Deviation
|
5.8
|
34.6
|
61.2
|
9.8
|
34.3
|
55.6
|
CV%
|
4.4
|
4.7
|
4.1
|
7.8
|
8.4
|
7
|
Cell Culture Supernates
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
107
|
604
|
1217
|
76.6
|
272
|
549
|
Standard Deviation
|
7.8
|
25
|
63.3
|
6.5
|
23.4
|
43.9
|
CV%
|
7.3
|
4.1
|
5.2
|
8.5
|
8.6
|
8
|
Recovery
The recovery of human IL-17 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=5)
|
97
|
88-112
|
Citrate Plasma (n=5)
|
105
|
92-116
|
EDTA Plasma (n=5)
|
101
|
90-110
|
Heparin Plasma (n=5)
|
101
|
93-109
|
Serum (n=5)
|
98
|
90-105
|
Linearity
To assess the linearity of the assay, five samples spiked with high concentrations of IL-17 were diluted with the appropriate Calibrator Diluent to produce samples with values within the dynamic range of the assay.