Human MIA Quantikine ELISA Kit
Human MIA Quantikine ELISA Kit
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human MIA Quantikine ELISA Kit
Catalog #
|
Availability
|
Size / Price
|
DMIA00
|
Contact your distributor for availability.
|
1 Kit /
|
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (25 uL), EDTA Plasma (25 uL), Heparin Plasma (25 uL), Saliva (50 uL)
Assay Range
15.6 - 1,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Saliva)
Specificity
Natural and recombinant human MIA
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human MIA Immunoassay is a 4.5 hour solid-phase ELISA designed to measure MIA in cell culture supernates, serum, plasma, and saliva. It contains E. coli-expressed recombinant Human MIA and antibodies raised against the recombinant factor. This immunoassay has been shown to accurately quantitate recombinant human MIA. Results obtained using natural human MIA showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural human MIA.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of a known concentration were tested in twenty separate
assays to assess inter-assay precision. Assays
were performed by at least three technicians using two lots of kit components
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Saliva
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
158
|
315
|
622
|
150
|
308
|
627
|
Standard Deviation
|
5.5
|
10.9
|
17.8
|
7.02
|
15.7
|
37.8
|
CV%
|
3.5
|
3.5
|
2.9
|
4.7
|
5.1
|
6
|
Recovery
The recovery of MIA spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=4)
|
103
|
101-107
|
EDTA Plasma (n=4)
|
98
|
85-115
|
Heparin Plasma (n=4)
|
99
|
85-113
|
Saliva (n=4)
|
102
|
86-110
|
Serum (n=4)
|
101
|
91-118
|
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of MIA were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay. Samples were diluted prior to assay.