Human MBL Quantikine ELISA Kit
Human MBL Quantikine ELISA Kit
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human MBL Quantikine ELISA Kit Summary
货号:DMBL00 规格:96T
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL)
Assay Range
0.2 - 10 ng/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Oligomeric forms of natural and recombinant human MBL
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human MBL immunoassay is a 4.5 hour solid phase ELISA designed to measure human MBL in cell culture supernates, serum, and plasma. It contains NS0-expressed recombinant human MBL and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human MBL showed dose-response curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values of naturally occurring human MBL.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
40
|
40
|
40
|
Mean
|
0.91
|
2.05
|
4.06
|
1.02
|
2.06
|
4.14
|
Standard Deviation
|
0.03
|
0.07
|
0.22
|
0.08
|
0.14
|
0.29
|
CV%
|
3.3
|
3.4
|
5.4
|
7.8
|
6.8
|
7
|
Recovery
The recovery of MBL spiked to various levels throughout the range of the assay was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Supernates (n=6)
|
113
|
106-117
|
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of natural MBL were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.