全国统一客服热线:
0755-28715175

产品中心
Product
4产品中心
Human LIF Quantikine ELISA Kit
Human LIF Quantikine ELISA Kit
Human LIF Quantikine ELISA Kit

Human LIF Quantikine ELISA Kit

 详情说明

Human LIF Quantikine ELISA Kit Summary

货号:DLF00

规格:96T

深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!

Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
3.5 hours or 4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (200 uL), Serum (200 uL), EDTA Plasma (200 uL), Heparin Plasma (200 uL)
Sensitivity
8 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human LIF
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.


Product Summary

The Quantikine Human LIF Immunoassay is a 3.5 or 4.5 hour solid phase ELISA designed to measure soluble human LIF in cell culture supernates, serum, and plasma. It contains E. coli-expressed recombinant human LIF and antibodies raised against the recombinant factor. This immunoassay has been shown to quantitate the recombinant LIF accurately. Results obtained measuring natural human LIF showed dose-response curves that were parallel to the standard curves obtained using the recombinant Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural human LIF.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision

Cell Culture Supernates

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean 341 875 1358 81.6 472 910
Standard Deviation 12.3 34.5 42.5 4.4 17.7 35.2
CV% 3.6 3.9 3.1 5.4 3.8 3.9

Serum, EDTA Plasma, Heparin Plasma

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean 314 868 1347 87.4 524 1017
Standard Deviation 7.7 24.3 23.8 6.2 34.9 46.5
CV% 2.5 2.8 1.8 7.1 6.7 4.6

Recovery

The recovery of LIF spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 97 91-106
EDTA Plasma (n=2) 89 84-97
Heparin Plasma (n=8) 92 84-100
Serum (n=11) 102 91-110


Copyright © 深圳9游总区生物科技有限公司版权所有

扫一扫,加关注

客户服务热线:
0755-33164177
13924667705
18665335907
0755-28715175