Human Chitinase 3-like 1 Quantikine ELISA Kit
Human Chitinase 3-like 1 Quantikine ELISA Kit
Human Chitinase 3-like 1 Quantikine ELISA Kit Summary
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DC3L10
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (10 uL), EDTA Plasma (10 uL), Heparin Plasma (10 uL), Urine (50 uL)
Assay Range
62.5 - 4,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine)
Specificity
Natural and recombinant human CHI3L1
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human CHI3L1 Immunoassay is a 4.5 hour solid phase ELISA designed to measure CHI3L1 in cell culture supernates, serum, plasma, and urine. It contains NS0-expressed recombinant human CHI3L1 and antibodies raised against recombinant human CHI3L1 and has been shown to accurately quantitate the recombinant factor. Results obtained using natural CHI3L1 showed dose response curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural human CHI3L1.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
40
|
40
|
40
|
Mean
|
566
|
1031
|
2202
|
606
|
1019
|
2212
|
Standard Deviation
|
26.7
|
44.4
|
102.6
|
32.1
|
59.1
|
153.1
|
CV%
|
4.7
|
4.3
|
4.7
|
5.3
|
5.8
|
6.9
|
Recovery
The recovery of CHI3L1 spiked to levels throughout the range of the assay was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=4)
|
97
|
93-100
|
Urine (n=4)
|
105
|
90-115
|
Linearity
To assess the linearity of the assay, samples containing high concentrations of CHI3L1 were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.