Mouse IL-17 Quantikine ELISA Kit
Mouse IL-17 Quantikine ELISA Kit
Mouse IL-17 Quantikine ELISA Kit Summary
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:M1700
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), EDTA Plasma (50 uL), Heparin Plasma (50 uL)
Assay Range
10.9 - 700 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant mouse IL-17
Cross-reactivity
Cross-reactivity observed with 1 or more available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Mouse IL-17 Immunoassay is a 4.5 hour solid phase ELISA designed to measure mouse IL-17 in cell culture supernates, serum, and plasma. It contains E. coli-expressed recombinant mouse IL-17 and antibodies raised against the recombinant factor. This immunoassay has been shown to quantitate recombinant mouse IL-17 accurately. Results obtained using natural mouse IL-17 showed linear curves that were parallel to the standard curves obtained using the recombinant Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for natural mouse IL-17.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
41.9
|
215
|
407
|
43.1
|
220
|
411
|
Standard Deviation
|
2.1
|
6.3
|
8.1
|
2.2
|
11.2
|
18.5
|
CV%
|
5
|
2.9
|
2
|
5.1
|
5.1
|
4.5
|
Recovery
The recovery of mouse IL-17 spiked to three levels throughout the range of the assay in various matrices was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Supernates (n=6)
|
101
|
92-118
|
EDTA Plasma (n=3)
|
107
|
93-120
|
Heparin Plasma (n=3)
|
101
|
91-123
|
Serum (n=6)
|
102
|
91-108
|
Linearity
To assess the linearity of the assay, five or more samples containing or spiked with various concentrations of mouse IL-17 in each matrix were diluted with Calibrator Diluent and then assayed.