Human CCL22/MDC Quantikine ELISA Kit
Human CCL22/MDC Quantikine ELISA Kit
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
Human CCL22/MDC Quantikine ELISA Kit Summary
货号:DMD00
Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL)
Assay Range
125.0 - 4,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)
Specificity
Natural and recombinant human MDC
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine Human MDC Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human MDC in cell culture supernates, serum, or plasma. It contains E. coli-expressed recombinant human MDC and has been shown to accurately quantitate the recombinant factor. Results obtained using naturally expressed human MDC showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally expressed human MDC.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision
Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
40
|
40
|
40
|
Mean
|
284
|
452
|
821
|
276
|
440
|
810
|
Standard Deviation
|
12
|
9.7
|
24
|
22.2
|
27.8
|
42.8
|
CV%
|
4.2
|
2.1
|
2.9
|
8
|
6.3
|
5.3
|
Recovery
The recovery of MDC spiked to three different levels throughout the range of the assay was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=4)
|
98
|
90-110
|
Linearity
To assess the linearity of the assay, samples containing or spiked with high concentrations of MDC were diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.\line