Human VEGF R1/Flt-1 Quantikine ELISA Kit
Human VEGF R1/Flt-1 Quantikine ELISA Kit
Human VEGF R1/Flt-1 Quantikine ELISA Kit Summary
深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!
货号:DVR100C
Sample Type & Volume Required Per Well
Cell Culture Supernates (100 uL), Cell Lysates (10 uL), Serum (100 uL), EDTA Plasma (100 uL), Heparin Plasma (100 uL), Saliva (100 uL)
Assay Range
31.3 - 2,000 pg/mL (Cell Culture Supernates, Cell Lysates, Serum, EDTA Plasma, Heparin Plasma, Saliva)
Specificity
Natural and recombinant human VEGF R1.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.
Product Summary
The Quantikine® Human VEGF R1/Flt-1 Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human VEGF R1 in cell culture supernates, cell lysates, serum, plasma, and saliva. It contains NS0-expressed recombinant human VEGF R1 and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human VEGF R1 showed linear curves that were parallel to the standard curves obtained using the Quantikine® kit standards. These results indicate that this kit can be used to determine relative mass values for natural human VEGF R1.
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components
Cell Culture Supernates, Cell Lysates, Serum, EDTA Plasma, Heparin Plasma, Saliva
|
Intra-Assay Precision
|
Inter-Assay Precision
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
Mean
|
207
|
610
|
1207
|
207
|
616
|
1205
|
Standard Deviation
|
5.16
|
14.9
|
35.6
|
13.9
|
38.9
|
74
|
CV%
|
2.5
|
2.4
|
2.9
|
6.7
|
6.3
|
6.1
|
Recovery
The recovery of human VEGF R1 spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type
|
Average % Recovery
|
Range %
|
Cell Culture Media (n=4)
|
99
|
87-110
|
Cell Lysis Buffer (n=4)
|
94
|
86-108
|
EDTA Plasma (n=4)
|
92
|
84-98
|
Heparin Plasma (n=4)
|
89
|
81-96
|
Saliva (n=4)
|
109
|
96-124
|
Serum (n=4)
|
104
|
93-114
|
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of human VEGF R1 were serially diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.