全国统一客服热线:
0755-28715175

产品中心
Product
4产品中心
Human Lipocalin-2/NGAL Quantikine ELISA Kit
Human Lipocalin-2/NGAL Quantikine ELISA Kit
Human Lipocalin-2/NGAL Quantikine ELISA Kit

Human Lipocalin-2/NGAL Quantikine ELISA Kit

 详情说明

Human Lipocalin-2/NGAL Quantikine ELISA Kit Summary

深圳9游总区生物优势代理美国R&DELISA系列产品试剂盒,价格优势,货期短,欢迎来电咨询0755-33164177!

货号:DLCN20

Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (50 uL), Heparin Plasma (50 uL), Saliva (50 uL), Urine (50 uL)
Sensitivity
0.04 ng/mL
Assay Range
0.2 - 10 ng/mL (Cell Culture Supernates, Serum, Heparin Plasma, Saliva, Urine)
Specificity
Natural and recombinant human Lipocalin-2
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.< 50% cross-species reactivity observed with species tested.
Interference
No significant interference observed with available related molecules.

Product Summary

The Quantikine Human Lipocalin-2 Immunoassay is a 4.5 hour solid-phase ELISA designed to measure human Lipocalin-2 in cell culture supernates, serum, plasma, urine, and saliva. It contains NS0-expressed recombinant human Lipocalin-2 and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human Lipocalin-2 showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring human Lipocalin-2.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision

Cell Culture Supernates, Serum, Heparin Plasma, Saliva, Urine

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 40 40 40
Mean 1.14 3.45 7.54 1.05 3.37 6.92
Standard Deviation 0.041 0.107 0.334 0.083 0.204 0.387
CV% 3.6 3.1 4.4 7.9 6.1 5.6

Recovery

The recovery of Lipocalin-2 spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 93 88-101
Urine (n=4) 90 84-97

Copyright © 深圳9游总区生物科技有限公司版权所有